Next Article in Journal
EPOPTIS: A Monitoring-as-a-Service Platform for Internet-of-Things Applications
Previous Article in Journal
UAV-Borne Mapping Algorithms for Low-Altitude and High-Speed Drone Applications
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Response of the TEROS 12 Soil Moisture Sensor under Different Soils and Variable Electrical Conductivity

by
Athanasios Fragkos
,
Dimitrios Loukatos
,
Georgios Kargas
* and
Konstantinos G. Arvanitis
Department of Natural Resources Management and Agricultural Engineering, Agricultural University of Athens, 75 Iera Odos Str., 11855 Athens, Greece
*
Author to whom correspondence should be addressed.
Submission received: 17 February 2024 / Revised: 27 March 2024 / Accepted: 28 March 2024 / Published: 29 March 2024
(This article belongs to the Topic Metrology-Assisted Production in Agriculture and Forestry)

Abstract

:
In this work, the performance of the TEROS 12 electromagnetic sensor, which measures volumetric soil water content (θ), bulk soil electrical conductivity (σb), and temperature, is examined for a number of different soils, different θ and different levels of the electrical conductivity of the soil solution (ECW) under laboratory conditions. For the above reason, a prototype device was developed including a low-cost microcontroller and suitable adaptation circuits for the aforementioned sensor. Six characteristic porous media were examined in a θ range from air drying to saturation, while four different solutions of increasing Electrical Conductivity (ECw) from 0.28 dS/m to approximately 10 dS/m were used in four of these porous media. It was found that TEROS 12 apparent dielectric permittivity (εa) readings were lower than that of Topp’s permittivity–water content relationship, especially at higher soil water content values in the coarse porous bodies. The differences are observed in sand (S), sandy loam (SL) and loam (L), at this order. The results suggested that the relationship between experimentally measured soil water content (θm) and εa0.5 was strongly linear (0.869 < R2 < 0.989), but the linearity of the relation θma0.5 decreases with the increase in bulk EC (σb) of the soil. The most accurate results were provided by the multipoint calibration method (CAL), as evaluated with the root mean square error (RMSE). Also, it was found that εa degrades substantially at values of σb less than 2.5 dS/m while εa returns to near 80 at higher values. Regarding the relation εab, it seems that it is strongly linear and that its slope depends on the pore water electrical conductivity (σp) and the soil type.

1. Introduction

The accurate estimation of volumetric soil water content (θ) helps in the efficient management of irrigation water to obtain high crop production but also in its low consumption In this regard, technologies like electronics, sensors, automation, networking and artificial intelligence have the potential to contribute to the transformation of the agri-food sector and provide systems capable of efficient processing, decision making, action and accurate real-time sensing [1,2,3,4]. The adaptation of irrigated agriculture to the new conditions is imperative in order to ensure that there will be sufficient quantities of water of suitable quality to support the functions of the natural environment both today and in the future.
Modern dielectric capacitance sensors, which are characterized by a low operating frequency (f), usually less than 100 MHz, provide the possibility of continuous recording of θ. Capacitance sensors determine apparent soil dielectric permittivity by measuring the charge time of a capacitor (i.e., the soil-probe system) for a given voltage. Due to a sensor’s low operating frequency, determination of the apparent dielectric permittivity is affected by bulk soil electrical conductivity (σb); as a result, the estimation of θ is not always accurate [5,6]. Thus, the accurate measurement of θ requires the special calibration of the device, i.e., the correlation of the (θ) and the apparent dielectric permittivity (εa) with the sensor measurements. Several of these sensors also measure other soil properties such as soil temperature (T) and bulk soil electrical conductivity (σb) [7,8].
Whereas many commercial sensors provide calibration equation, which combine the square root of the apparent dielectric permittivity with the soil water content, several researchers report the possibility of further improving this specific manufacturer equation [9,10]. Toward this direction, the main objective of this study was to evaluate the accuracy of the empirical calibration equations for the TEROS 12 sensor, as given by the manufacturer. An important question is whether or not the primary (i.e., uncalibrated) measurements, as provided by the sensor for different soil types and for different salinity conditions, give the expected values, when put into the manufacturer calibration equations, for (θ) and (εa). The detailed and independent evaluation of this sensor aims to provide fertile ground for improving its overall performance toward a better exploitation of its potential. A prototype device, utilizing widely available electronics was developed for facilitating the raw measurement process.

2. Background Work and Motivations

Almost all soil sensors available today use εα as a basis for determining θ. This approach is based on the highly apparent dielectric permittivity of water (εα = 80) as opposed to soil solids (εα = 2–5) or air (εα = 1). Topp et al. [11] demonstrated that dielectric measurements could be used to accurately measure the (θ) using the TDR (Time Domain Reflectometry) method. The large difference between the values of the apparent dielectric permittivity of water and the other constituents of soil makes the apparent dielectric permittivity sensitive to θ.
The TDR method is able to accurately calculate a material’s dielectric permittivity from wave propagation, because of the fact that a material’s permittivity and its water content are strongly related, as it was also presented by Hoekstra and Delaney [12] and Topp et al. [11]. By measuring electromagnetic pulse reflection time, we calculate θ indirectly by determining the apparent dielectric permittivity. The time required for a high-frequency electromagnetic wave (300–1000 MHz) to traverse the length of a waveguide is measured in order to assess the εα, which is directly related to the response time of the wave. Although TDR is a reliable method for verifying soil or substrate θ, its expensive cost has led in the development of alternative dielectric sensors technologies (capacitance, impedance, and transmission line oscillators), which are cheaper and do not rely on complicated waveform analysis. The capacitance sensors determine the charging time of the sensor capacitor depending on apparent dielectric permittivity. The sensor TEROS 12 uses the enclosing soil surrounding the three needles as the dielectric of the capacitor, from which it measures its charge time and output a raw voltage based on the substrate εa.
The dielectric methods have been extensively studied and constitute an accepted measurement technique as it does not disorder the ground and is suitable for automation and remote sensing. Furthermore, the improvement of θ estimation has been attempted, taking into account the effect of (σb) on apparent dielectric permittivity [7,13,14,15,16].
The response of a dielectric material to electromagnetic fields is expressed by a complex function [17].
ε* = ε′ − jε″,
of which the real part (ε′) expresses the storage of electric energy in the medium while the imaginary part (ε″) expresses its diffusion (loss factor) in the means of propagation, with j = 1 . If we divide the relation (1) by the dielectric constant of the vacuum εο = 8.85∙10−12 Fm−1, we obtain its conversion to the relation (2)
ε*r = ε′r − jε″r,
where ε*r is relative apparent permittivity, the real part ε′r is the dielectric constant that expresses how much energy can be stored by the electromagnetic field imposed and is mainly related to soil moisture θ as well as to the spatial distribution of the soil’s liquid phase while the imaginary part ε″r expresses the dielectric losses that occur when the mechanism of polarization of water in the soil does not follow exactly the changes of the applied field. Polarization is the rearrangement of charges inside the material, to compensate the electric field (i.e., the positive and negative charges as they move in different directions). These relations are critical because dielectric permittivity is directly related to both ε′r and ε″r for almost all soil water sensors.
The dielectric constant and dielectric losses vary with applied frequency f and their detailed study should focus on the frequency range operated by a soil sensor. The real part of the relative apparent permittivity appears to take the largest values at low operation frequencies and generally decreases with increasing frequency.
Other soil factors that affect apparent permittivity are in general: pore and specific surface area distributions, grain and pore shape and size, organic matter content, fabric and structure of soil, porosity, chemical composition of the soil solution, temperature, and bulk soil electrical conductivity (σb) [18,19,20,21]. Due to the number of these parameters we focused to evaluate the effects of (σb) on the sensor performance and estimate if a specific calibration procedure is needed.
For the case of TDR, it has been shown that dielectric permittivity can be related to θ with reasonable accuracy, and thus θ can be calculated from it, for a wide variety of soils, using a single calibration equation [11]. Alternatives to empirically derived calibrations are often based on the dielectric mixing model. Ferre and Topp [22] indicated that soil water content can be calculated from εa using a simple regression formula that relates θ to εa, using the form of Equation (3), where a and b are suitable fitting parameters.
θ = a ε a + b
In this regard, for the TDR measurements, Topp and Reynolds [23], using the general form of Equation (3) with parameter α equal to 0.115 and parameter b equal to −0.176, have shown that the derived formula, Equation (4),
θ = 0.115 ε a 0.176
is effectively equivalent to the third-order calibration equation suggested by Topp et al. in 1980 [11] and it deviates less than 0.01 m3 m−3 from it, for θ values calculated over a range from 0.05 to 0.45 m3 m−3.
The existence of a corresponding linear relationship θ-ε0.5 has been demonstrated for many dielectric sensors. But there is a big difference in the values of parameters a, b (Equation (3)) compared to those values for the case of TDR (Equation (4)) [7,24,25,26,27,28]. These values are affected by the type of soil, the operating frequency of the sensor, the temperature, the bulk soil electrical conductivity and the special characteristics of the sensor. Nasta et. al., [8] examined two soils (loamy and clay) and showed that the application of soil-specific temperature correction is highly recommended, if the TEROS 12 is installed in regions with pronounced seasonal variations in soil temperature.
However, the question remains open whether a linear relationship θ-ε0.5 is valid in many types of soil and whether the coefficients of the linear relationship differ depending on the type of soil. We also investigate whether the accuracy of the sensors is affected and how by changes in soil salinity.

3. Materials and Methods

3.1. Soil Sensor Characteristics

The TEROS 12 sensor from METER Group AG indirectly measures the volumetric water content in the soil as well as the apparent dielectric permittivity of the porous medium by emitting an electromagnetic wave frequency of 70 MHz, and also measures temperature using a thermistor which is in the center needle and bulk electrical conductivity. It has three stainless steel needles (i.e., instead of the two that the TEROS 10 sensor has), each 5.5 cm long, and is compatible with most data recording systems. The sensor delivers an electromagnetic wave to the needles that are charged depending on the dielectricity of the material. The charging time is proportional to the dielectric current and θ of the substrate. The resulting measurement volume is approximately 1010 cm3. The TEROS 12 sensor comes pre-calibrated for mineral soil and has the following specifications volumetric water ranges from 0 to 0.70 m3 m−3, accuracy is ±3% in mineral soils with standard factory calibration when the electrical conductivity of the solution is EC < 8 dS/m, the temperature measurement range is −40 to 60 °C with a resolution of 0.1 °C and the measurement accuracy is ±1 °C. According to its manual [29], the electrical conductivity measurement range of TEROS 12 is 0–10 dS/m, with a resolution of 0.001 dS/m and a measuring accuracy of 5%. The TEROS 12 sensor is known for its low energy consumption (3.5 mA in the operative mode and 0.03 mA in the sleep mode).
According to the TEROS 12 manual for third-party data loggers, each fresh raw θ sensor value (RAW) should be put into the proper calibration equation and the corresponding value be extracted. The equations for calibrating and deriving the real values from the raw values are given by the following equations, for soil moisture:
θ (m3 m−3) = 3.879 ∙ 10−4 ∙ RAW − 0.6956,
and for the dielectric permittivity:
εa = (2.887 ∙ 10−9 ∙ RAW3 − 2.080 ∙ 10−5 ∙ RAW2 + 5.276 ∙ 10−2 ∙ RAW − 43.39)2,
As a remark, for the bulk conductivity provided by the sensor, the original raw values were following two different formats (i.e., in dS/m and in μS/cm) according to their magnitude, and thus a division of the raw values by 1000 was necessary for the small quantities (expressed in μS/cm), a practice that its necessity was also observed by Tadaomi Saito et al. [30].

3.2. Technical Arrangements for Data Collection and Processing

In order to exploit the data provided by the TEROS 12 soil sensor, a cost-effective microcontroller was properly wired and programmed, based on the ATmega32U4 chip. This microcontroller is a 3.3 V logic unit and is equipped with analog input pins and digital input/output pins. It also has a USB port for easy connectivity and programming purposes via a conventional computer/laptop. For the latter task, the Arduino IDE environment has been the preferred option, which also provides fast and easy debugging and monitoring options via its Serial Monitor and the Serial Plotter components.
Prior to interfacing the soil instrument with the prototype measuring equipment, a thorough study of the sensor physiology was necessary. More specifically, according to the manufacturer’s instructions for TEROS 12, a voltage of 4 to 15 V DC should be supplied while its input can tolerate signals from 2.8 to 5.0 V. The digital output of this sensor is compatible with the 3.3 V logic. The physical connection with the microcontroller is addressed via a custom measuring circuit able to host the 3.5 mm stereo plug connector the soil sensor has. Special care should be taken as the TEROS 12 in its default settings, after warming up, transmits a fresh triplet of soil volumetric water content (VWC), temperature and bulk conductivity values according to the simple serial protocol called DDI; and, after this, it enters the more complex serial communication mode called SDI-12. For taking the measurements that this research is using, the DDI method is adequate, and thus the necessary code was written in the microcontroller memory, according to the timing diagram provided by the sensor manufacturer. A voltage converter raised the voltage from 3.3 to 5 Volts to supply the sensor, and a multiplexer directed the data to the desired serial input line, in case of more than one instrument connected with the same acquisition unit. The microcontroller was also responsible for periodically powering on and off the sensor, every 10 s. Technical arrangement details supporting the data acquisition process are shown in Figure 1.
The RAW data acquired from the TEROS 12 via the microcontroller were applied in the calibration equations (implemented on the microcontroller) to calculate the soil moisture and the dielectrical permittivity according to Equations (3) and (4), respectively. The latter values, along with the original triplet (i.e., the VWC, temperature and the bulk conductivity values) were forwarded to a laptop, connected to the microcontroller via a USB connection, for storage and further processing and/or visualization, typically utilizing the MS Excel application. Sensor-to-sensor variability effects were ignored in our work by crudely assuming limited sensor-to-sensor variability, similar to Nasta et al. in [8].

3.3. Measurement in Soils

The experiments were conducted using six different inorganic soil types. In greater detail, 5 of the 6 soil samples being mentioned were collected from agricultural areas and more specifically from the Prefecture of Argolida [31]. The 6th one, i.e., the sand sample, is used as part of artificial substrates in floriculture. All experiments were performed at constant temperature (23 ± 1 °C) to avoid temperature effects. All the air-dried soils were ground sieved to reach particle size ≤ 2 mm and were later put in the oven to calculate the initial air-dried soil moisture, via measuring their weight loss. Calculation of the dry bulk density of every sample was performed to convert the moisture from a mass base to a volume base. The corresponding soil properties are presented in Table 1. These soils were chosen because they exhibit a wide variety of properties. As can be seen from Table 1, the clay content ranges from 0% to 48%. However, the study of the characteristics of the pore structure and the type of clay in each soil is under investigation [32]. Soils samples of sand, clay, loam, sandy loam, sandy clay loam, and silty clay loam were used, in which an increase in θ took place in a strictly defined manner up to the saturation point. In particular, an air-dried soil sample volume of 1500 mL was used, to which 75 mL of tap water (EC = 0.28 dS/m) was added each time in order to increase the actual water content (θm) each time by θ = 0.05 m3 m−3. Pressed with a 0.20 kg hammer, the soil material of different water content levels was aimed to achieve uniform density distribution in the soil sample. The θm and dry bulk density of the soils were determined again by weighing and oven drying at the end of the experiment. In addition, measurements were taken in an oven-dried soil sample (θ = 0 m3 m−3). Via this step sequence, a sufficient number of εa, σb and θ values for all the soil types examined were provided. For each value of θm, four measurements of the RAW and T, σb data acquired from the TEROS 12 via the microcontroller were applied in the calibration equations (implemented on the microcontroller) and then the average was calculated. This was considered to be appropriate, with the scope of detecting any errors during the measurements.
At the same time, it was ensured that the walls of the container are more than 3 cm apart, which the manufacturer requires at least in order to have no effect on the measurements and not affect the readings of the sensor. The sensor was vertically inserted into the soil samples and measurements were taken. Figure 2 depicts characteristic instances from the soil preparation and measurement process via the TEROS 12 sensor.

3.4. Salinity Effects in Water Solutions and Soils

To investigate the effect of salinity on the performance of TEROS 12, experiments with increased ECw were carried out in aqueous solutions and in soils. Initially, measurements in water solutions with increasing electrical conductivity values (ECw) from 0 to 20 dS/m by adding potassium salt (i.e., potassium chloride—KCl) to deionized water were made in order to investigate TEROS 12 sensitivity and response to ECw changes. The sensor was inserted vertically into water solutions and the output readings were determined in different ECw. This experiment is a guide for the effect of ECw on the stability of apparent dielectric permittivity measurements.
Moreover, the experimental procedure mentioned in the part «3.3 Measurements in soils» was used in order to evaluate the sensor’s sensitivity to salinity for four soil materials (sand, sandy loam, loam, clay). In addition to the KCl solutions of ECw = 0.28 dS/m, similar experiments were carried out with KCl solutions of ECw 3, 6 and 10 dS/m. For sandy clay loam and silty clay loam the response of the TEROS 12 was tested and evaluated only for ECw = 0.28 dS/m.

3.5. Prediction of Pore-Water Electrical Conductivity

As a final step in this study, the prediction of soil pore water electrical conductivity (σp) by utilizing TEROS 12 sensor measurements has been attempted. For this purpose the σp prediction model, of the Hilhorst [33] model was used. This model is connecting εa, σb and σp through the equation
  σ p = ε p   σ b ε a ε 0
where εp is the apparent dielectric permittivity of the pore water, considered equal to that of pure water (80.3 at 20 °C), and ε0 is the y-intercept of the line εa = f(σb). Using Equation (7), σp can be calculated from a simultaneous measurement of εa and σb.
Hilhorst [33] found that ε0 was in the range of 1.9–7.6. The value of 4.1 was proposed to be used as a mean value for all soil types.

3.6. Performance Evaluation Criteria

To evaluate the efficiency of the various calibration equations, the root mean square error (RMSE) and the correlation coefficient R2 were used as evaluation criteria.
R M S E = 1 n i = 1 n ( O i P i ) 2
R 2 = i = 1 n ( O i P i ) 2 i = 1 n ( O i O ¯ ) 2
where O is the observed value, O ¯ is the mean observed value, P is the predicted value of soil water content, i is the counter for data pairs and n is the total number of different pairs of observed–predicted values. Good prediction requires data with low bias and error, gibing a low root mean square error. When RMSE approaches 0, this means we have very good prediction, while when it approaches 1, we have a prediction about the same as the average value. The correlation coefficient R2 varies between 0 and 1; when it approaches 0, there is no correlation, while we have a very strong correlation when it approaches 1. In intermediate values between 0 and 1, we have a weak, mean or strong correlation.

4. Results and Discussion

4.1. Salinity Effects in Liquid Solutions

In Figure 3, the sensor (εa) response to increasing ECw in water is shown, where ECw = σb is the complex. The εa values, which should be between 78 and 80, decrease at approximately the value of 60 with increasing salinity to the value of 2.5 dS/m, stabilizing at approximately the value of 60 between the values of 2.5 and 6 dS/m, then increasing up to 80 with increasing salinity to the value of 10 dS/m and then remaining constant at approximately 80. This behavior is contrary to what is reported by Seyfried and Murdock [25] for the Hydra Probe sensor, by Hilhorst [30] for the Sigma probe sensor, by Kargas et al. [6] for the WET sensor and Kargas and by Soulis [7] for the CS655 sensor, which operate at low frequencies and are not affected by the increase in salinity up to 3 dS/m. As observed, there is a high sensitivity of the TEROS 12 to relatively low σb values. This behavior is contrary to what was expected due to the effect of the low operating frequency of TEROS 12 and may only be attributed to the sensor design. Rosenbaum et al. [34], Schwartz et al. [15] and Kargas et al. [6] reported similar sensitivity abnormality studying the 5TE sensor behavior.

4.2. Soil-Specific Calibration

To better study the TEROS 12 behavior, further experiments were conducted using soil samples instead of liquid solutions. The corresponding results are depicted in Figure 4. In greater detail, the relationship between the apparent dielectric permittivity (unitless) and the actual soil water content, i.e., εam, was determined (as depicted by blue points in Figure 4a–f—EXP) for characteristic soil samples and for a salinity level of ECw = 0.28 dS/m. The soil samples were sand, sandy loam, loam, clay, silty clay loam and sandy clay loam. For better comparison purposes, two other curves are also depicted in each soil case figure. The first curve (black line) expresses the relationship εa-θ where θ is calculated by Equation (4) (TOPP curve) and the second curve (red line) expresses the relationship εa-θ where the θ is calculated by the soil-specific CAL calibration equation (CAL curve).
By inspecting the results in Figure 4, it is inferred that TEROS 12 always gives lower values for εa, compared to the ones derived via the Topp equation for the same θm level. The greatest difference is observed in sand, i.e., 12 units. In sandy loam, the difference becomes slightly lower, i.e., 10.5 units, while it further drops to 5 units in loam. In clay, it is observed that the values of εa, given by TEROS 12, are closer to the Topp values.
These results are reflected in the calibration parameters shown in Table 2, where both a and b parameters are substantially greater than the Topp equation ones for all soil types being used. These results are quite unexpected because the TEROS 12 f is only 70 MHz, much smaller than that for TDR. As was shown from Table 2, the θm0.5 relationship is strongly linear in all soils. Additionally, it appears that each soil type has a separate calibration equation. Thus, a single calibration equation could not work for all soil types. The corresponding R2 values as well as the σb readings obtained by the TEROS 12 sensor at saturation are also presented in Table 2. More specifically, the R2 values range from 0.952 to 0.989.
The corresponding RMSE values for the θ predictions obtained by the specific calibration method and the manufacturer calibration are presented in Table 3.
From the results, it is observed that the CAL procedure has much smaller RMSE values compared to the manufacturer calibration equation. More specifically, the RMSE values for the CAL calibration method range from 0.013 to 0.046 m3 m−3, while the RMSE values for the manufacturer calibration range from 0.027 to 0.098 m3 m−3. Using CAL parameters, the average RMSE was reduced by almost 50%. These results indicate that the CAL is an effective soil-specific procedure.

4.3. Salinity Effects in Soils

The experimental arrangements described in Section 3.4 were applied in order to investigate the sensor’s response and sensitivity to the indicative salinity levels of 0.28, 3, 6 and 10 dS/m and for four soils, namely sand, sandy loam, loam and clay. In this case, KCl solutions with known ECw were added to obtain predetermined θm levels from dry to saturation.
As obvious from Figure 5a, when ECw increases, the εa value for the same θm decreases instead of increasing, because of the low-frequency operation of TEROS 12.
The greatest values of the coefficients of the equation θm0.5 in sand are observed in the large values of bulk electrical conductivity σb because the values εa decrease with the increase of σb (max σb sand = 1.870 dS/m) (Table 2). This behavior is similar to the corresponding in the solutions with increasing salinity level. It appears that the unexpected behavior of the TEROS 12 sensor was related to its response to σb, which was highly nonlinear and inverse, from negative to small σb to positive to large σb. Tadaomi Saito et al. [30] reported a similar behavior of the TEROS 12 sensor.
On the contrary, in the clay soil (Figure 5d), which is most affected by ε″, the calibration relationships are closer to the Topp equation than the sand especially in the higher ECw value. In the case of clay soil, the decrease in εa may be compensated from its increase because of the effect of ε″. Consequently, the coefficients of the calibration equations have a lower value compared to those of sand.
Sandy loam and loam soils (Figure 5b and Figure 5c, respectively) exhibit an intermediate behavior. More specifically, up to a θm value of approximately 0.3 m3 m−3, the values of εa corresponding to ECw = 0.28 dS/m are higher than those corresponding to ECw = 10 dS/m, while the tendency reverses above this value of θm.
However, even though this complication appears, a strong linearity of the θm0.5 relationship is again observed with increasing σb. As can be seen from Table 2, the value of R2 is always greater than 0.925. Only in the case of loam at ECw = 10 dS/m, the R2 value is relatively small (R2 = 0.869).
Nevertheless, it is demonstrated that the linearity decreases with the increase in σb values. The most characteristic case is loam, where R2 = 0.961 in ECw = 0.28 dS/m, whereas R2 = 0.869 in ECw = 10 dS/m.

4.4. Prediction of Pore-Water Electrical Conductivity

Τhe εa measurements (for different ECw levels and for each of the characteristic soil samples) were plotted against the corresponding σb measurements. More specifically, experiments involved ECw levels equal to 0.28, 3, 6, 10 dS/m, for the sand, sandy loam, loam and clay soil types. Furthermore, two additional experiment sets were conducted for silty clay loam and for sandy clay loam, with a salinity level of ECw = 0.28 dS/m. The resulting curves are shown in Figure 6a–f, respectively.
As can be clearly seen, in Figure 6, the slope of the line is salinity dependent and decreases when the ECw increases. In sand at ECw = 0.28 dS/m, the slope of the line is 19.360, whereas at EC = 10 dS/m the corresponding slope is approximately three times lower (5.729). Also, it seems that the value of the slope is affected by the soil type. Concerning the value of ε0, it appears that is much lower than the 4.1 value with a higher value, 3.1, for sand, sandy loam and clay soils. In the loam, silty clay loam and sandy clay loam soil types, the values of ε0 are approximately 4.
The obtained results using the TEROS 12 sensor revealed that the εab relationship is strongly linear in all salinity level cases, since the values of R2 ranged between 0.878 and 0.997, as shown in Table 4.
From Table 5, it appears that the Hilhorst model predicts values of σp considerably higher than those of the wetting solution ECw values. However, if the values of the electrical conductivity of the extract’s saturation paste (ECe), which were measured for each soil (Table 5), are taken into account, the soils initially contained relatively high amounts of salts.
Considering that the initial values of σp are approximately twice the values of ECe in the field capacity, the overestimation of σp is not that great. More specifically in the loam soil for ECw = 3 dS/m, the value of σp = 6.21 dS/m. This value is approximately equal to the sum of ECw plus twice the value of ECe. Consequently, the Hilhorst model combined with the TEROS 12 sensor satisfactorily approximates the salinity level in soils.

5. Conclusions

The performance of the TEROS 12 soil moisture sensor was evaluated under six different soil types and four salinity levels under laboratory experiments. The behavior of the TEROS 12 sensor is complex as the EC increases. In liquids with increasing EC, there is an anomalous behavior of the TEROS 12 sensor. This behavior is related to its response to σb, which is highly nonlinear and in the reverse direction, from negative at small σb to positive at large σb.
The θm- ε relationship was strongly linear in all examined cases but linearity was reduced at the high salinity level. The CAL calibration procedure, as evaluated with the RMSE, was effective for all soils. Additionally, it appears that each soil type has a separate calibration equation. The above underlines the importance of carrying out a specific laboratory experiment for the correct conversion of apparent dielectric permittivity to soil water content (θ), especially under a regime of salinity fluctuations. In these cases, the application of salinity-independent factory calibration, especially in coarse soils, leads to significant errors in the calculation of θ.
The εab relationship sensor is strongly linear for all salinity levels. The slope of the linear relationship is salinity dependent and is also affected by the soil type. The Hilhorst model is approximately adequate in measuring σp in soils with the TEROS 12 sensor.

Author Contributions

Conceptualization, A.F., D.L. and G.K; methodology, A.F., D.L. and G.K; prototyping, D.L.; software, D.L.; validation, A.F., D.L. and G.K.; investigation, A.F. and D.L.; data curation, A.F. and D.L.; writing—original draft preparation, A.F., D.L. and G.K.; writing—review and editing, A.F., D.L., G.K. and K.G.A.; visualization, A.F. and D.L.; supervision, G.K. and K.G.A. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Available upon request.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Liu, Y.; Ma, X.; Shu, L.; Hancke, G.P.; Abu-Mahfouz, A.M. From industry 4.0 to agriculture 4.0: Current status, enabling technologies, and research challenges. IEEE Trans. Ind. Informat. 2020, 17, 4322–4334. [Google Scholar] [CrossRef]
  2. Javaid, M.; Haleem, A.; Singh, R.P.; Suman, R. Enhancing smart farming through the applications of Agriculture 4.0 technologies. Int. J. Intell. Netw. 2022, 3, 150–164. [Google Scholar] [CrossRef]
  3. Chen, Y.; Or, D. Geometrical factors and interfacial processes affecting complex dielectric permittivity of partially saturated porous media. Water Resour. Res. 2006, 42, W06423. [Google Scholar] [CrossRef]
  4. Rosenzweig, C.; Elliott, J.; Deryng, D.; Ruane, A.C.; Müller, C.; Arneth, A.; Boote, K.J.; Folberth, C.; Glotter, M.; Khabarov, N.; et al. Assessing agricultural risks of climate change in the 21st century in a global gridded crop model intercomparison. Proc. Natl. Acad. Sci. USA 2014, 111, 3268–3273. [Google Scholar] [CrossRef] [PubMed]
  5. Regalado, C.; Ritter, A.; Rodríguez-González, R. Performance of the Commercial WET Capacitance Sensor as Compared with Time Domain Reflectometry in Volcanic Soils. Vadose Zone J. 2007, 6, 244–254. [Google Scholar] [CrossRef]
  6. Kargas, G.; Κerkides, P.; Seyfried, M.S. Response of Three Soil Water Sensors to Variable Solution Electrical Conductivity in Different Soils. Vadose Zone J. 2014, 13, 1–13. [Google Scholar] [CrossRef]
  7. Kargas, G.; Soulis, K. Performance evaluation of a recently developed soil water content, dielectric permittivity, and bulk electrical conductivity electromagnetic sensor. Agric. Water Μanagement 2019, 213, 568–579. [Google Scholar] [CrossRef]
  8. Nasta, P.; Coccia, F.; Lazzaro, U.; Bogena, H.R.; Huisman, J.A.; Sica, B.; Mazzitelli, C.; Vereecken, H.; Romano, N. Temperature-Corrected Calibration of GS3 and TEROS-12 Soil Water Content Sensors. Sensors 2024, 24, 952. [Google Scholar] [CrossRef] [PubMed]
  9. Karmacharya, D.A. Large-Scale Experiment on Performance of MIDP in Stratified Silty Sand, Arizona State University ProQuest Dissertations Publishing,  2023. Available online: https://hdl.handle.net/2286/R.2.N.187403 (accessed on 20 January 2024).
  10. Peranić, J.; Čeh, N.; Arbanas, Ζ. The Use of Soil Moisture and Pore-Water Pressure Sensors for the Interpretation of Landslide Behavior in Small-Scale Physical Models. Sensors 2022, 22, 7337. [Google Scholar] [CrossRef]
  11. Topp, G.C.; Davis, J.L.; Annan, A.P. Electromagnetic determination of soil water content: Measurements in coaxial transmission lines. Water Resourses Res. 1980, 16, 574–582. [Google Scholar] [CrossRef]
  12. Hoekstra, P.; Delaney, A. Dielectric properties of soils at UHF and microwave frequencies. J. Geophys. Res. 1974, 79, 1699–1708. [Google Scholar] [CrossRef]
  13. Wyseure, G.; Mojid, M.A.; Manzoor, A.M. Measurement of volumetric water content by TDR in saline soils. Eur. J. Soil Sci. 2005, 48, 347–354. [Google Scholar] [CrossRef]
  14. Robinson, D.A.; Jones, S.B.; Wraith, J.M.; Or, D.; Friedman, S.P. A Review of Advances in Dielectric and Electrical Conductivity Measurement in Soils Using Time Domain Reflectometry. Vadose Zone J. 2003, 2, 444–475. [Google Scholar] [CrossRef]
  15. Schwartz, R.C.; Casanova, J.J.; Pelletier, M.G.; Evett, S.R.; Baumhardt, R.L. Soil permittivity response to bulk electrical conductivity for selected soil water sensors. Vadose Zone J. 2013, 12, 1–13. [Google Scholar] [CrossRef]
  16. Mohamed, A.-M.O.; Paleologos, E.K. Chapter 16—Dielectric Permittivity and Moisture Content. In Fundamentals of Geoenvironmental Engineering; Mohamed, A.-M.O., Paleologos, E.K., Eds.; Butterworth-Heinemann: Oxford, UK, 2018; pp. 581–637. [Google Scholar] [CrossRef]
  17. Von Hippel, A. Dielectrics and Waves; John Wiley: Hoboken, NJ, USA, 1953; Available online: https://is.muni.cz/el/1431/podzim2015/F7061/um/Dielectric_and_Waves.pdf (accessed on 20 January 2024).
  18. Rhoades, J.D.; Ratts, P.; Prather, R. Effects of liquid-phase electrical conductivity, water content, and surface conductivity on bulk soil electrical conductivity. Soil Sci. Soc. Am. J. 1976, 40, 651–655. [Google Scholar] [CrossRef]
  19. Rhoades, J.D.; Manteghi, N.A.; Shouse, P.J.; Alves, W.J. Soil electrical conductivity and soil salinity: New formulations and calibrations. Soil Sci. Soc. Am. J. 1989, 53, 433–439. [Google Scholar] [CrossRef]
  20. Mualem, Y.; Friedman, S. Theoretical prediction of electrical conductivity in saturated and unsaturated soil. Water Resour. Res. 1991, 27, 2771–2777. [Google Scholar] [CrossRef]
  21. Wobschall, D. A frequency shift dielectric soil moisture sensor. IEEE Trans. Geosci. Electron. 1978, 16, 112–118. [Google Scholar] [CrossRef]
  22. Topp, G.C.; Ferre, P.A. The Soil Solution Phase. Methods of Soil Analysis: Part 4; Wiley: Hoboken, NJ, USA, 2018; pp. 417–545. [Google Scholar] [CrossRef]
  23. Topp, G.C.; Reynolds, W.D. Time domain reflectometry: A seminal technique for measuring mass and energy in soil. Soil Tillage Res. 1998, 47, 125–132. [Google Scholar] [CrossRef]
  24. Kargas, G.; Kerkides, P. Performance of the theta probe ML2 in the presence of nonuniform soil water profiles. Soil Tillage Res. 2009, 103, 425–432. [Google Scholar] [CrossRef]
  25. Seyfried, M.S.; Murdock, M.D. Measurement of soil content with a 50 MHz soil dielectric sensor. Soil Sci. Soc. Am. J. 2004, 68, 394–403. [Google Scholar] [CrossRef]
  26. Kargas, G.; Soulis, K. Performance analysis and calibration of a new low-cost capacitance soil moisture sensor. J. Irrig. Drain. Eng. 2011, 138, 632–641. [Google Scholar] [CrossRef]
  27. Arsoy, S.; Ozgur, M.; Keskin, E.; Yilmaz, C. Enhancing TDR based water content measurements by ANN in sandy soils. Geoderma 2013, 195–196, 133–144. [Google Scholar] [CrossRef]
  28. Bogena, H.R.; Huisman, J.A.; Schilling, B.; Weuthen, A.; Vereecken, H. Effective calibration of low-cost soil water content sensors. Sensors 2017, 17, 208. [Google Scholar] [CrossRef] [PubMed]
  29. Meter. User guide for TEROS 11/12. Available online: http://publications.metergroup.com/Manuals/20587_TEROS11-12_Manual_Web.pdf (accessed on 22 January 2024).
  30. Saito, T.; Oishi, T.; Inoue, M.; Iida, S.; Mihota, N.; Yamada, A.; Shimizu, K.; Inumochi, S.; Inosako, K. Low-Error Soil Moisture Sensor Employing Spatial Frequency Domain Transmissometry. Sensors 2022, 22, 8658. [Google Scholar] [CrossRef] [PubMed]
  31. Kargas, G.; Chatzigiakoumis, I.; Kollias, A.; Spiliotis, D.; Massas, I.; Kerkides, P. Soil Salinity Assessment Using Saturated Paste and Mass Soil:Water 1:1 and 1:5 Ratios Extracts. Water 2018, 10, 1589. [Google Scholar] [CrossRef]
  32. Chen, X.; Luo, Y.; Shao, L.; Niu, G. Three-dimensional pore structure characteristics of granite residual soil and their relationship with hydraulic properties under different particle gradation by X-ray computed tomography. J. Hydrol. 2023, 618, 129230. [Google Scholar] [CrossRef]
  33. Hilhorst, M.A. A pore water conductivity sensor. Soil Sci. Soc. Am. J. 2000, 64, 1922–1925. [Google Scholar] [CrossRef]
  34. Rosenbaum, U.; Huisman, J.A.; Vrba, J.; Vereecken, H.; Bogena, H.R. Correction of temperature and electrical conductivity effects on dielectric permittivity measurements with ECH2O sensors. Vadose Zone J. 2011, 10, 582–593. [Google Scholar] [CrossRef]
Figure 1. Technical arrangement details supporting the data acquisition process: (a) hardware; (b) software.
Figure 1. Technical arrangement details supporting the data acquisition process: (a) hardware; (b) software.
Sensors 24 02206 g001
Figure 2. Characteristic instances from the soil preparation and measurement process.
Figure 2. Characteristic instances from the soil preparation and measurement process.
Sensors 24 02206 g002
Figure 3. The εa values against ECw in aqueous KCl solutions for the TEROS 12 and the 5TE sensors.
Figure 3. The εa values against ECw in aqueous KCl solutions for the TEROS 12 and the 5TE sensors.
Sensors 24 02206 g003
Figure 4. The apparent dielectric permittivity (unitless) and soil water content (in cm3 cm−3) relationship (i.e., εam) for characteristic soil samples and for a salinity of ECw = 0.28 dS/m (blue points). More specifically, the soil cases were: (a) sand; (b) sandy loam; (c) loam; (d) clay; (e) silty clay loam; (f) sandy clay loam. Two other curves are also depicted per soil type. The first curve (black line) expresses the relationship εa-θ, where θ is calculated by Equation (4) (TOPP curve) and the second curve (red line) expresses the relationship εa-θ, where the θ is calculated by soil-specific CAL calibration equation (CAL curve).
Figure 4. The apparent dielectric permittivity (unitless) and soil water content (in cm3 cm−3) relationship (i.e., εam) for characteristic soil samples and for a salinity of ECw = 0.28 dS/m (blue points). More specifically, the soil cases were: (a) sand; (b) sandy loam; (c) loam; (d) clay; (e) silty clay loam; (f) sandy clay loam. Two other curves are also depicted per soil type. The first curve (black line) expresses the relationship εa-θ, where θ is calculated by Equation (4) (TOPP curve) and the second curve (red line) expresses the relationship εa-θ, where the θ is calculated by soil-specific CAL calibration equation (CAL curve).
Sensors 24 02206 g004aSensors 24 02206 g004b
Figure 5. The apparent dielectric permittivity (unitless) and soil water content (in cm3 cm−3) relationship (i.e., εam) for specific salinity levels and for the characteristic soil samples: (a) sand; (b) sandy loam; (c) loam; (d) clay.
Figure 5. The apparent dielectric permittivity (unitless) and soil water content (in cm3 cm−3) relationship (i.e., εam) for specific salinity levels and for the characteristic soil samples: (a) sand; (b) sandy loam; (c) loam; (d) clay.
Sensors 24 02206 g005
Figure 6. Τhe relationship εab for various ECw levels and for the characteristic soil types: (a) sand; (b) sandy loam; (c) loam; (d) clay; (e) silty clay loam; (f) sandy clay loam. For the last two types of soil, the relation εab refers only to ECw = 0.28 dSm−1.
Figure 6. Τhe relationship εab for various ECw levels and for the characteristic soil types: (a) sand; (b) sandy loam; (c) loam; (d) clay; (e) silty clay loam; (f) sandy clay loam. For the last two types of soil, the relation εab refers only to ECw = 0.28 dSm−1.
Sensors 24 02206 g006
Table 1. Soil Properties.
Table 1. Soil Properties.
Soil TypeClaySiltSandDry Bulk Density
%%%(g/cm3)
Sand 1001.66 ± 0.010
Sandy Loam1611731.24 ± 0.010
Loam1932491.23 ± 0.011
Clay4812401.13 ± 0.012
Sandy Clay Loam2512631.26 ± 0.011
Silty Clay Loam3149201.11 ± 0.011
Table 2. Parameters a and b in the relationship θ-ε0.5 for all soil types and salinity levels along with R2 values, respectively, and maximum σb values at saturation.
Table 2. Parameters a and b in the relationship θ-ε0.5 for all soil types and salinity levels along with R2 values, respectively, and maximum σb values at saturation.
Soil TypeEC (dS/m)baR2Max σb (dS/m)
Sand0.28−0.2560.1710.9730.580
3−0.2880.1950.9560.940
6−0.2580.1940.9251.300
10−0.2400.1860.9381.870
Sandy Loam0.28−0.2260.1460.9800.563
3−0.2070.1440.9541.130
6−0.2470.1650.9671.420
10−0.1810.1350.9522.140
Loam0.28−0.2090.1420.9610.530
3−0.2020.1260.9890.930
6−0.1820.1320.9741.720
10−0.1520.1290.8692.130
Clay0.28−0.2750.1430.9521.590
3−0.2030.1470.9322.120
6−0.2100.1210.9392.550
10−0.1770.1090.9252.460
Silty Clay Loam0.28−0.2990.1720.9880.440
Sandy Clay Loam0.28−0.2760.1550.9540.800
Table 3. The RMSE of θ relative to those calculated using manufacturer calibration TEROS 12 and those derived from the CAL procedure.
Table 3. The RMSE of θ relative to those calculated using manufacturer calibration TEROS 12 and those derived from the CAL procedure.
Manufacturer CalibrationCAL
Soil TypeSalinity LevelRMSEAverageRMSEAverage
Sand0.280.0560.0820.0190.026
30.0810.024
60.0910.031
100.0980.031
Sandy Loam0.280.0270.0420.0140.023
30.0430.027
60.0570.023
100.0410.028
Loam0.280.0400.0480.0250.029
30.0530.026
60.0330.020
100.0650.046
Clay0.280.0430.0470.0330.032
30.0500.030
60.0470.031
100.0480.035
Silty Clay Loam0.280.044 0.013
Sandy Clay Loam0.280.040 0.027
Table 4. Parameters a and b in the relationship εa-σb for all soil types and salinity levels along with R2 values, respectively.
Table 4. Parameters a and b in the relationship εa-σb for all soil types and salinity levels along with R2 values, respectively.
Soil TypeECBaR2
Sand0.281.16619.3600.985
32.3169.3440.989
62.1266.6040.987
101.9745.7290.997
Sandy Loam0.281.29123.0180.963
32.76111.9380.985
62.1048.6000.990
101.4807.6250.998
Loam0.283.78425.4850.997
34.56012.7860.993
64.1308.9650.990
104.0167.6800.992
Clay0.282.29612.2680.921
33.10010.3360.986
61.8809.7980.929
100.68310.3690.878
Silty Clay Loam0.284.23325.7400.989
Sandy Clay Loam0.283.93215.4620.988
Table 5. Soil types, values of the conductivity extract’s saturation paste (ECe) and values of the electrical conductivity of the pore water (σp) as they have been calculated by the Hilhorst (2000) model.
Table 5. Soil types, values of the conductivity extract’s saturation paste (ECe) and values of the electrical conductivity of the pore water (σp) as they have been calculated by the Hilhorst (2000) model.
Soil TypeEC (dS/m)σp (dS/m)ECe (dS/m)
Sand0.285.672.770
310.32
615.48
1017.17
Sandy Loam0.284.541.387
37.40
610.87
1012.30
Loam0.283.371.880
36.21
68.54
1010.01
Clay0.286.341.807
37.72
68.30
107.26
Silty Clay Loam0.284.912.120
Sandy Clay Loam0.282.921.570
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Fragkos, A.; Loukatos, D.; Kargas, G.; Arvanitis, K.G. Response of the TEROS 12 Soil Moisture Sensor under Different Soils and Variable Electrical Conductivity. Sensors 2024, 24, 2206. https://0-doi-org.brum.beds.ac.uk/10.3390/s24072206

AMA Style

Fragkos A, Loukatos D, Kargas G, Arvanitis KG. Response of the TEROS 12 Soil Moisture Sensor under Different Soils and Variable Electrical Conductivity. Sensors. 2024; 24(7):2206. https://0-doi-org.brum.beds.ac.uk/10.3390/s24072206

Chicago/Turabian Style

Fragkos, Athanasios, Dimitrios Loukatos, Georgios Kargas, and Konstantinos G. Arvanitis. 2024. "Response of the TEROS 12 Soil Moisture Sensor under Different Soils and Variable Electrical Conductivity" Sensors 24, no. 7: 2206. https://0-doi-org.brum.beds.ac.uk/10.3390/s24072206

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop